Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0356920120630020157
Korean Journal of Anesthesiology
2012 Volume.63 No. 2 p.157 ~ p.160
Wire-reinforced endotracheal tube fire during tracheostomy -A case report-
Shin Young-Duck

Lim Seung-Woon
Bae Jin-Ho
Yim Kyoung-Hoon
Sim Jae-Hwan
Kwon Eun-Jung
Abstract
Every operation could have a fire emergency, especially in the case of a tracheostomy. When a flammable gas meets a source of heat, the danger of fire is remarkable. A tracheal tube filled with a high concentration of oxygen is also a great risk factor for fire. Intra-tracheal tube fire is a rare, yet critical emergency with catastrophic consequences. Thus, numerous precautions are taken during a tracheostomy like, use of a special tube to prevent laser damage, ballooning of the tube with normal saline instead of air, and dilution of FiO2 with helium or nitrogen. Since the first recorded cases on tube fires, most of the fires were initiated in the balloon and the tip. In the present case report, however, we came across a fire incidence, which originated from the wire.
KEYWORD
Airway management, Burns, Complications, Tracheostomy
FullTexts / Linksout information
Listed journal information
MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø