Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0356920130640050407
Korean Journal of Anesthesiology
2013 Volume.64 No. 5 p.407 ~ p.413
Impact of priming the infusion system on the performance of target-controlled infusion of remifentanil
Kim Jong-Yeop

Moon Bong-Ki
Lee Jong-Hyuk
Min Sang-Kee
Jo Youn-Yi
Abstract
Background: The start-up behavior of syringe and syringe pump is known to be one of the causes of inaccurate intravenous infusion. This study evaluated the method of priming the infusion system (PRIMING), and its impact on the target-controlled infusion (TCI) of two remifentanil diluents.

Methods: PRIMING was performed using an evacuation of 2.0 ml to the atmosphere prior to TCI. Forty-eight TCI, using 50 ¥ìg/ml (Remi50) or 20 ¥ìg/ml (Remi20) of diluents, were performed targeting 4.0 ng/ml of effect-site concentration (Ceff), with PRIMING or not. The gravimetrical measurements of the delivered infusates reproduced actual Ceff. The bolus amount and time to reach 95% target were compared.

Results: Without PRIMING, Remi50 infused less bolus (43 ¡¾ 23 %) than Remi20 (19 ¡¾ 9 %) (P = 0.003), and showed more delayed increase of Ceff (11.2 ¡¾ 4.0 min) than Remi20 (7.4 ¡¾ 0.4 min) (P = 0.028). However, PRIMING significantly decreased the deficit of the bolus (2 ¡¾ 1%), as well as the delay of the increase of Ceff in Remi50 (1.2 ¡¾ 0.2 min) (both P < 0.001). In addition, with PRIMING, the start-up bolus showed minimal difference to the nominal bolus (1 and 2%), and Ceff were increased to 4.0 ¡¾ 0.1 ng/ml at the expected time of peak effect, irrespective of the diluents.

Conclusions: Proper operation of the syringe pump used in the priming of the syringe may be helpful in reduction of the inaccuracy of TCI, particularly during the early phase of infusion, or the infusion of a more concentrated diluent.
KEYWORD
Drug delivery, Priming, Target-controlled infusion
FullTexts / Linksout information
Listed journal information
MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø