Unidirectional valve malfunction by the breakage or malposition of disc - two cases report -
|
|
Lee Chol
Lee Kyu-Chang Kim Hye-Young Kim Mi-Na Choi Eun-Kyung Kim Ji-Sub Lee Won-Sang Lee Myeong-Jong Kim Hyung-Tae
|
|
Abstract
|
|
|
Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.
|
|
KEYWORD
|
|
Capnography, Hypercapnia, Unidirectional valve
|
|