KMID : 0357520060290010013
|
|
Journal of Radiological Science and Technology 2006 Volume.29 No. 1 p.13 ~ p.19
|
|
A Study of Uniformity Test in PET/CT
|
|
Kim Su-Geun
Jeong Hyi-Il Park Soung-Ock
|
|
Abstract
|
|
|
The PET scanner can detect the photon pair arriving from the source in phantom. The number of light photons released by the crystal(scintillator-BGO or LSO). In recent scintillation crystals in block structures were incorporated into full ring systems, and their resulted marked improvement in spatial resolution and increase in a sensitivity to annihilations. The uniformity of the crystal sensitivity is very important to makes correct information of abnormal states in organs. These factors influenced by the dection efficiency of the scintillators. We have study about the uniformity of crystals to the annihilation, And study about the standard deviation to average counts. The relative standard deviation in central detector groups more uniformed than circumferenced detector groups. It is caused detected quanta of gamma ray by the geometrical factors of PET detector. PET cameras are available with different geometric arrangement and several parallel rings oriented in the axial direction. The center groups from 7th to 40th groups are comparatively uniform and sensitive. But at the circumferenced detectors decreased the sensitivity and uniformity.
|
|
KEYWORD
|
|
detection efficiency, standard deviation, average count
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|