Journal of Radiological Science and Technology
2007 Volume.30 No. 2 p.113 ~ p.119
Evaluation of MTF Image by Target/Filter Combined of X-ray Tube Using Mammography
Yang Han-Jun

Ko Shin-Gwan
Ko Shin-Gwan
Joo Mi-Hwa
Abstract
It is important to consider the contrast of object in Mammography because an absorption gap between tissues of body and breast in breast is low. This study is to evaluate MTF image with resolution chart according to change of combination of target and filter. The results were as follows : 1. There were significant differences in X-ray energy according to combination of filter(Mo/Mo, Mo/Rh. Mo/Al, Rh/Rh, Rh/Al) and acrylic thickness(2 cm, 3 cm, 4 cm). 2. The value of lp/mm on MTF to 0.5 showed that the sharpness in MTF curve was 2.4 compared to Mo/Mo and 2cm acryl, 2.63 in Mo/Rh and 4 cm acryl, and 2.9 in Rh/Rh and 6cm acryl. 3. The value of lp/mm on MTF showed that the resolution in MTF curve was 6.0 compared to Mo/Mo and 2 cm acryl, 4.60 in Rh/Al and 4cm acryl, and 6.03 in Rh/Al and 6 cm acryl. 4. The value of MTF on 2.5 lp/mm distinguishable visually was 0.48 compared to Mo/Mo and 2 cm acryl, 0.53 in Mo/Rh and 4cm acryl, and 0.59 in Rh/Rh and 6cm acryl. 5. For the evaluation of an image of the mammo-phantom, the score of Mo/Mo was 12 points, Mo/Rh 11, Rh/Rh 10.5, Mo/Al 10, Rh/Al 9.0, respectively.
KEYWORD
Target/Filter combination, MTF, Mammographic phantom