Journal of Radiological Science and Technology 2007 Volume.30 No. 2 p.139 ~ p.145
|
|
Evaluation of Dose Distribution Using Gafchromic BBT(R) Film
|
|
Kang Se-Sik
Ko Seong-Jin Ko Seung-Jin Jang Eun-Sung
|
|
Abstract
|
|
|
Dose evaluation for small field such as stereotactic radiosurgery was performed using EBT film. Every film which irradiated 6MV photon beam was scanned and obtained the optical density(OD) by flat bed scanner after 24 hours of irradiation. This study compared dose from diode in water and Gafchromic film in acrylic phantom to verify the reliability of the film, and to evaluate the SRS in clinical dose distributions from calculation and measurement in the region of virtual target in humanoid and cylindrical phantoms were compared. The Gafchromic film was found to be linear up to 9Gy. The for 6 MV was measured at 1.5 cm from the surface by both of diode and the film. As the depth is deeper, the error was measured within at depth. Comparing between distribution from calculation and measurement, we found that there is 5% error at 90% isodose line. We found that given dose could be measured accurately by using the phantoms. It was feasible to use the Gafchromic film in quality assurance of SRS.
|
|
KEYWORD
|
|
radiosurgery, Gafchromic BBT(R) film, phantoms, quality assurance
|
|