Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520070300030245
Journal of Radiological Science and Technology
2007 Volume.30 No. 3 p.245 ~ p.250
Measurement of Apron Shielding Rate for X-ray and Gamma-ray
Park Myeong-Hwan

Kwon Deok-Moon
Abstract
This research measured the shielding rates of apron 0.25 and 0.5 mmPb for X-ray energy in diagnosis radiation system and gamma-ray energy of -MDP and -FDG. X-ray energies were measured on effective energy of when additional filtering plate of 0, 2 mmAl is used within the range of tube voltage , and at this time, apron 0.5 mmPb has shown about 5.5% of increase in its shielding rate over 0.25 mmPb at the highest quality. Besides, the aprons of the two types have shown high shielding rate of over 90% for direct X-ray and spatial dose rate. And, in case 0.25 and 0.5 mmPb aprons were used at 140keV of -MDP, the shielding effects were between 30 and 53%, and at high energy of 511 keV, -FDG, the shielding effects of apron, , were very small.
KEYWORD
Apron, Shielding rate, Effective energy, Spatial dose rate
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)