Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520080310010089
Journal of Radiological Science and Technology
2008 Volume.31 No. 1 p.89 ~ p.95
Determination of Quality Correction Factors for a Plane-Parallel Chamber in High Energy Electron Beams using Monte Carlo Calculation
Jeong Dong-Hyeok

Lee Jeong-Ok
Abstract
The quality correction factor for used beam and qualities is strongly required for clinical dosimetry by TRS-398 protocol of IAEA. In this study the quality correction factors for a commercial plane-parallel ionization chamber in high energy electron beams were calculated by Monte Carlo code(DOSRZnrc/EGSnrc). In comparison of quality correction factor, the difference between this study and TRS-398 were within 1% in 5-20 MeV. In case of 4MeV the difference was 1.9%. As an independent method of determination of quality correction factor this study can be applied to evaluate values in the protocol or calculate the factor for a new chamber.
KEYWORD
Quality Correction Factor, Electron Dosimetry, Monte Carlo
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)