KMID : 0357520090320030261
Journal of Radiological Science and Technology
2009 Volume.32 No. 3 p.261 ~ p.270
Relationship of Compressed Breast Thickness and Average Glandular Dose According to Focus/Filter
Lee In-Ja

Abstract
The study examined the relationship between the compressed breast thickness and Average Glandular Dose (AGD) among 1,969 outpatients who went through breast X-ray in a university hospital for 10 months from July 1st, 2007 to April 30th, 2008. Then it analyzed the result acquired from 3,900 cases of Cranio-Caudal (CC) view, especially, when the breasts were compressed (13-15daN). The following is the conclusion driven from the relationship analysis. 1. The subjects aged in 40s and 50s were 2,679 out of 3,900 cases and this figure was 68.69% in all. 2. In terms of distribution depending on focus/filter, 41.0% was Mo/Mo, 34.8% was Mo/Rh, and 24.2% was Rh/Rh. 3. In terms of compressed breast thickness depending on focus/filter, the average thickness was 26.91 mm at Mo/Mo, 38.84 mm at Mo/Rh, and 48.80 mm at Rh/Rh. The average thickness of the entire cases was shown to be 36.27 mm. 4. AGD depending on focus/filter was 1.27 mGy at Mo/Mo, 1.55 mGy at Mo/Rh, and 1.42 mGy at Rh/Rh. The average glandular dose of the entire cases was shown to be 1.43 mGy. 5. The relationship of AGD depending on compressed breast thickness at Mo/Mo was y=0.0318x + 0.470 while it was y=0.0206x + 0.709 at Mo/Rh and y=0.0248x + 0.335 at Mo/Rh. It was highly influenced by the compressed breast thickness, however, more variation was detected at Mo/Mo depending on breast thickness.
KEYWORD
Average Glandular Dose, Compressed Breast Thickness, Focus/Filter, Mo/Mo, Mo/Rh, Mo/Rh
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)