KMID : 0357520090320030307
Journal of Radiological Science and Technology
2009 Volume.32 No. 3 p.307 ~ p.312
Effectiveness of Bismuth Shield to Reduce Eye Lens Radiation Dose Using the Photoluminescence Dosimetry in Computed Tomography
Jung Mi-Young

Kweon Dae-Cheol
Gwon Soo-Il
Abstract
The purpose of our study was to determine the eyeradiation dose when performing routine multi-detector computed tomography (MDCT). We also evaluated dose reduction and the effect on image quality of using a bismuth eye shield when performing head MDCT. Examinations were performed with a 64MDCT scanner. To compare the shielded/unshielded lens dose, the examination was performed with and without bismuth shielding in anthropomorphic phantom. To determine the average lens radiation dose, we imaged an anthropomorphic phantom into which calibrated photoluminescence glass dosimeter (PLD) were placed to measure the dose to lens. The phantom was imaged using the same protocol. Radiation doses to the lens with and without the lensshielding were measured and compared using the Student t test. In the qualitative evaluation of the MDCT scans, all were considered to be of diagnostic quality. We did not see any differences in quality between the shielded and unshielded brain. The mean radiation doses to the eyewith the shield and to those without the shield were 21.54 versus 10.46 mGy, respectively. The lens shield enabled a 51.3% decrease in radiation dose to the lens. Bismuth in-plane shielding for routine eye and head MDCT decreased radiation dose to the lenswithout qualitative changes in image quality. The other radiosensitive superficial organs specifically must be protected with shielding.
KEYWORD
Bismuth, CT, Dose, Protection
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)