Journal of Radiological Science and Technology
2009 Volume.32 No. 4 p.393 ~ p.399
A Study on the Factors of Spatial Scattered Ray Occurrence in the X-ray Radiography Room
Na Soo-Kyung

Han Sang-Hyo
Abstract
In this study, we measured the dose distribution of scattered ray in X-ray radiography room using an ion chamber and examined the dependency of scattered ray content on the scattered ray source and exposure condition. To study the factors of scattered ray occurrence in the acryl phantom, we measured the change in the scatted ray content according to the X-ray tube voltage (40~140 kV) and the field size (, , ). For the field size, the side-scattering rate ranged from 3.1% to 14.5%. The scattered ray contributions of the phantom, collimator, X-ray tube and wall were also measured. The scattered ray contribution of the phantom was higher than 95.4% for the entire tube voltage, and those of the collimator, X-ray tube and wall were 2.6%, 1.3% and 0.7%, respectively.
KEYWORD
spatial distribution, scattered ray, field size, contribution