KMID : 0357520100330010025
|
|
Journal of Radiological Science and Technology 2010 Volume.33 No. 1 p.25 ~ p.29
|
|
Development and Usefulness of Acrylic Needle for Percutaneous Bone Biopsy
|
|
Kim Tae-Hyung
Shin Myung-Jin Shin Ji-Hoon Lim Jin-Oh Ryu Ji-Yeon Oh Jae-Seon Woo Chul-Woong Nam Jeung-Hee
|
|
Abstract
|
|
|
To evaluate the usefulness of newly designed percutaneous bone biopsy needle for pumpkin¡¯s specimen collection. We manufactured three types of biopsy needle with different internal processing which were made of 10 mm-diameter acrylic material. We made the conventional type (Ct) similar to the clinical type then compared the test group. The type a (Ta) made 1 cm-length internal processing from the distal, type b (Tb) made taper, and type c (Tc) made internal processing like spiral configuration. We performed 20 times biopsy to get an 10 mm length specimen from pumpkin¡¯s surface and evaluated the success rate of the biopsy, length of the specimen, and determine internal processing type of the most suitable biopsy needle (ANOVA test). Success rates of Ct, Ta, Tb, and Tc were 55%, 80%, 90%, and 100%, respectively. The lengths of the specimen of Ct, Ta, Tb, and Tc were , , , and , respectively. All groups were statistically significant (p < 0.05) except the group between Ct and Ta (p = 0.28). Newly designed bone biopsy needle seems to be useful for obtaining enough specimen. Tc may be more effective than other types.
|
|
KEYWORD
|
|
Percutaneous Bone Biopsy, Bone Biopsy Needle
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|