KMID : 0357520100330020085
Journal of Radiological Science and Technology
2010 Volume.33 No. 2 p.85 ~ p.91
A Quality Assurance on Digital Chest Radiography in Medical Institution for Pneumoconiosis : Compared with Analog Radiography
Lee Won-Jeong

Ko Kyung-Sun
Park Jai-Soung
Kim Sung-Jin
Chu Sang-Deok
Park So-Young
Choi Byung-Soon
Abstract
Digital radiography has been replacing rapidly the analog radiography for diagnosis of pneumoconiosis. The purpose of this study is to compare quality control of digital radiography (DR) and analog radiography (AR) for chest radiography in medical institution for pneumoconiosis (MIP) For the first time, we visited MIP to evaluate the chest radiography which is used for patients with pneumoconiosis, including equipment, technical parameters and reading environment. There were 33 institutions. DR and AR were installed in 24 and 9 institutions, respectively. Between DR and AR, we compared the radiological technique (RT), image quality (IQ) and reading environment (RE) to use the guideline published by Occupational Safety and Health Research Institute (OSHRI). The image quality was rated by two experienced chest radiologists for pneumoconiosis with certified from OSHRI. The chest radiography equipment was not significantly difference between AR and DR, but there were significantly difference in tube voltage and grid ratio used for chest radiography except to tube current, exposure time. Statistically, DR is significantly higher in RT(70.3 vs. 43.8, p = 0.009), RE(77.7 vs. 33.3, p = 0.004) than AR, but it¡¯s not significantly difference in IQ (65.6 vs. 52.8, p = 0.050). AR and DR in RT were passed 33.3%, 75.0% respectively (p = 0.044) and 44.4%, 79.2% (p = 0.090) in IQ and 44.4%, 91.7% (p = 0.009) in RE. In MIP, DR needs to replace AR in diagnosis of pneumoconiosis.
KEYWORD
Medical institution for pneumoconiosis, Quality assurance, Digital radiography, Analog radiography
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)