Journal of Radiological Science and Technology
2010 Volume.33 No. 3 p.173 ~ p.178
Evaluation of the Patient Dose in Case of Standard Radiographic Examinations Using CR and DR
Kim Sang-Tae

Han Beom-Hui
Abstract
In projection radiography, two types of digital imaging systems are currently available, computed radiography (CR) and digital radiography (DR): a difference between them can be stated in terms of dose and image quality. In the Department of Radiology our hospital, a flat-panel DR equipment (Digital diagnost, Philips) and two CR systems (ADC Compact plus digitizer, AGFA) are employed. Eight standard radiographic examinations (Skull AP, Skull LAT, Chest PA, Chest LAT, Abdomen AP, L-spine AP, L-spine LAT, Pelvis AP) were considered: doses delivered to patients in terms of both entrance skin dose (ESD) were calculated and compared in order to study the dosimetric discrepancies between CR and DR. Assessment of image quality is undertaken by Consultant Radiologists to ensure that the quality criteria for diagnostic radiographic images of the European guidelines were met. Results showed that both ESD in DR are lower than that in CR; all images met the criteria in the European Guidelines for both modalities and were used for reporting by the radiologists. Since the operators are the same and the image quality is comparable in both modalities, this study shows that in the considered examinations, DR can perform better than CR from a dosimetric point of view.
KEYWORD
Computed radiography, Digital radiography, Entrance skin dose