KMID : 0357520100330030277
Journal of Radiological Science and Technology
2010 Volume.33 No. 3 p.277 ~ p.282
Verification of Non-Uniform Dose Distribution in Field-In-Field Technique for Breast Tangential Irradiation
Park Byung-Moon

Bae Yong-Ki
Kang Min-Young
Bang Dong-Wan
Kim Yon-Lae
Lee Jeong-Woo
Abstract
The study is to verify non-uniform dose distribution in Field-In-Field (FIF) technique using two-dimensional ionization chamber (MatriXX, Wellhofer Dosimetrie, Germany) for breast tangential irradiation. The MatriXX and an inverse planning system (Eclipse, ver 6.5, Varian, Palo Alto, USA) were used. Hybrid plans were made from the original twenty patients plans. To verify the non-uniform dose distribution in FIF technique, each portal prescribed doses (90 cGy) was delivered to the MatriXX. The measured doses on the MatriXX were compared to the planned doses. The quantitative analyses were done with a commercial analyzing tool (OmniPro IMRT, ver. 1.4, Wellhofer Dosimetrie, Germany). The delivered doses at the normalization points were different to average 1.6% between the calculated and the measured. In analysis of line profiles, there were some differences of 1.3-5.5% (Avg: 2.4%), 0.9-3.9% (Avg: 2.5%) in longitudinal and transverse planes respectively. For the gamma index (criteria: 3 mm, 3%) analyses, there were shown that 90.23-99.69% (avg: 95.11%, std: 2.81) for acceptable range (-index 1) through the twenty patients cases. In conclusion, through our study, we have confirmed the availability of the FIF technique by comparing the calculated with the measured using MatriXX. In the future, various clinical applications of the FIF techniques would be good trials for better treatment results.
KEYWORD
Field-In-Field (FIF), 2D Array Ionization chamber (MatriXX), -index
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)