Journal of Radiological Science and Technology
2011 Volume.34 No. 4 p.351 ~ p.357
Analysis of Relationship Between Injection Dose and Exposure Dose in PET/CT Scan: Initial Study
Park Hoon-Hee

Lyu Kwang-Yeul
Abstract
The 18F-FDG is one of the widely used isotopes for PET/CT scans. Dose amount injected to the patient depends on the characteristics of PET/CT systems. Obviously, the technologists who contact with patients would be exposed as well. In this study, we evaluated the exposed dose of the technologist who works on the PET/CT scanner. The exposed dose were measured every month with the TLDs from 6 technologists. Each technologist is shift-worker who manages 3 different PET/CT systems(Scanner 1(S1): 0.15 mCi/kg, Scanner 2(S2): 0.17 mCi/kg, Scanner 3(S3): 0.12 mCi/kg). The average exposed doses of technologists for each PET/CT system were measured as 0.76 mSv for S1, 0.93 mSv for S2 and 0.47 mSv for S3. The maximum dose was 1.12 mSv and minimum was 0.42 mSv. The results showed that there was a correlation between exposed dose and PET/CT system(p<0.005). Less injected dose for patient occurs less exposed dose for technologist. Various studies for the low dose PET/CT system are required for not only the patient but also the technologist.
KEYWORD
PET/CT, Exposure Dose, Injection Dose