KMID : 0357520120350030219
|
|
Journal of Radiological Science and Technology 2012 Volume.35 No. 3 p.219 ~ p.226
|
|
A Study on Radiation Exposure Dose of Operator During Interventional Radiology Procedure
|
|
Jeon Mi-Eun
Lim Cheong-Hwan Jung Hong-Ryang You In-Gyu Hong Dong-Hee Kang Byung-Sam
|
|
Abstract
|
|
|
Interventional radiologists are not aware of the potential injury from procedures. The purpose of this study are to evaluate radiation exposure of interventional radiologist from intervention procedures and to develop guideline of the simple methods for decreasing their radiation exposure from intervention procedure. In this study, Dosimeters were used to monitor operator doses of radiation exposure in a broad range of procedures from 20 interventional radiologist during the periods of 3 months. And, we searched protection methods of each interventional radiologist. During TACE procedure, there was using 0.5 mmPb radiation protector decreased average 89.5 % of radiation exposure. Thicker radiation protector provide decreasing radiation exposure. And radiation exposure dose decreased average 47.7 % by using pulse fluoroscopic mode. Therefore, interventional radiologist should wear protective aprons, use active shielding, monitor their doses, and know how to poisoning themselves during the procedure and operate correct of the machines for minimum dose.
|
|
KEYWORD
|
|
Interventional Radiology, Radiation Exposure Dose, Radiation Protection
|
|