Journal of Radiological Science and Technology
2012 Volume.35 No. 4 p.321 ~ p.326
The Impact of Signal Intensity and Image Distortion Magnetic Resonance Imaging in the Orthopedic Prosthetic Metal
Kim Hyeong-Gyun

Choi Seong-Dae
Abstract
Used as an ingredient in the hospital for orthopedic prosthetic stainless and titanium metal the same size as on the MRI diagnostic value of imaging were compared. Center of images, background high band portion of the video signal is converted into a weighted intensity values Normal images and compared. The area of normal slice and also the distortion of images and cross-sectional imaging of a range of quantitative and sagittal planes were compared. As a result, the periphery high band signal intensity values of Stainless video phantom 2, pig bone 1.8, Titanium imaging of phantom 1.7 has higher value than Normal video pig bone 1.3 times the signal strength rivers. MRI distortion of the shape and the distortions of the range, if the cross-sectional area compared to Normal Slice Stainless case of phantom 65.8 %, pig bone 61.5 %, Titanium distortion phantom 23.1 %, pig bone 38.5 % of the range of community found. In this experiment, as a result, MRI was found to be Titanium more diagnostic value than the specimen with respect to the signal intensity weighted value and low distortion range, Stainless.
KEYWORD
MRI, Orthopedic, Image Distortion, Stainless, Titanium