Journal of Radiological Science and Technology
2013 Volume.36 No. 1 p.11 ~ p.17
The Usefulness of Al Face Block Fabrication for Reducing Exposure Dose of Thyroid Glands in Mammography
Hong Eun-Ae

Lee In-Ja
Abstract
Currently, there are many studies being conducted around the world to reduce exposure dose to radiation for patients to receive medical treatments in a safe environment. We developed and fabricated of this shield that the patients are protected from the radiation and are need of safety control during breast imaging. In this study, for breast imaging, GE Senography 2000D were used and set at SID 65cm, 28kVp, and 63mAs. The measuring instrument was Flukes Victoreen 6000-529. And we performed Face Block on with 30 patients. The chamber on the actual thyroid glands to take CC and MLO and measure the dosage before and after wearing the Face Block. For the results, after wearing the Face Block, exposure was decreased by 53.8%-100% and 65.8% in average in CC View and by 50%-100% and 60.7% in average in MLO View. The development of the Face Block that practically decreased the exposure dose of thyroid glands, crystalline eyes during breast imaging and reduced the patients anxiety during breast imaging. The Face Block is expected to improve patients satisfaction and contribute to reducing patients exposure dose, but more efforts should be made to reduce exposure dose to medical radiation.
KEYWORD
Mammography, Exposure Dose, Face Block