KMID : 0357520140370010001
Journal of Radiological Science and Technology
2014 Volume.37 No. 1 p.1 ~ p.6
Testing and Analysis of Tube Voltage and Tube Current in The Radiation Generator for Mammography
Jung Hong-Ryang

Hong Dong-Hee
Han Beom-Hui
Abstract
Breast shooting performance management and quality control of the generator is applied to the amount of current IEC(International Electrotechnical Commission) 60601-2-45 tube voltage and tube current are based on standards that were proposed in the analysis of the test results were as follows. Tube voltage according to the value of the standard deviation by year of manufacture from 2001 to 2010 as a 42-3.15 showed the most significant, according to the year of manufacture by tube amperage value of the standard deviation to 6.38 in the pre-2000 showed the most significant, manufactured after 2011 the standard deviation of the devices, the PAE(Percent Average Error) was relatively low. This latest generation device was manufactured in the breast of the tube voltage and tube diagnosed shooting the correct amount of current to maintain the performance that can be seen. The results of this study as the basis for radiography diagnosed breast caused by using the devices performance and maintain quality control, so the current Food and Drug Administration "about the safety of diagnostic radiation generator rule" specified in the test cycle during three years of self-inspection radiation on a radiation generating device ensure safety and performance of the device using a coherent X-ray(constancy) by two ultimately able to keep the radiation dose to the public to reduce the expected effect is expected.
KEYWORD
The Radiation Generator for Mammography, IEC, Tube voltage, Tube Current
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)