KMID : 0357520140370020135
Journal of Radiological Science and Technology
2014 Volume.37 No. 2 p.135 ~ p.141
A Study on Characteristic of Glass Dosimeter According to Graded Change of Tube Current
Son Jin-Hyun

Kim Seong-Ho
Mun Hyun-Jun
Kim Lyun-Kyun
Son In-Hwa
Kim Young-Jun
Min Jung-Whan
Kim Ki-Won
Abstract
This study was evaluated the linearity and reproducibility according to dose, and reproducibility according to delay time by changing tube current amount (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively, which are low energy radiations) using Glass Dosimeter (GD) and piranha semiconductor dosimeter which are used for measuring exposure dose. Measurements of radiation dose were performed using external detector of piranha 657 which is multi-function QA device (RTI Electronic, Sweden). Conditions of measurement were 80 kVp, SSD 100 cm and exposure region is$10cm{ imes}10cm$ ¼ö½Ä À̹ÌÁö . Glass dosimeter was exposed to radiation. Twenty-four glass dosimeters were divided into six groups (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively), then measured. This study was resulted by measuring the linearity and reproducibility according to change of tube current in low energy field. In dose characteristic of GD, this study could be useful as previous study with regard to dose characteristic according to change of tube voltage in low energy field.
KEYWORD
Thermoluminescence Dosimeter (TLD), Glass Dosimeter (GD), Linearity, Reproducibility