A Study on Characteristic of Glass Dosimeter According to Graded Change of Tube Current
|
|
Son Jin-Hyun
Kim Seong-Ho Mun Hyun-Jun Kim Lyun-Kyun Son In-Hwa Kim Young-Jun Min Jung-Whan Kim Ki-Won
|
|
Abstract
|
|
|
This study was evaluated the linearity and reproducibility according to dose, and reproducibility according to delay time by changing tube current amount (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively, which are low energy radiations) using Glass Dosimeter (GD) and piranha semiconductor dosimeter which are used for measuring exposure dose. Measurements of radiation dose were performed using external detector of piranha 657 which is multi-function QA device (RTI Electronic, Sweden). Conditions of measurement were 80 kVp, SSD 100 cm and exposure region is$10cm{ imes}10cm$ ¼ö½Ä À̹ÌÁö . Glass dosimeter was exposed to radiation. Twenty-four glass dosimeters were divided into six groups (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively), then measured. This study was resulted by measuring the linearity and reproducibility according to change of tube current in low energy field. In dose characteristic of GD, this study could be useful as previous study with regard to dose characteristic according to change of tube voltage in low energy field.
|
|
KEYWORD
|
|
Thermoluminescence Dosimeter (TLD), Glass Dosimeter (GD), Linearity, Reproducibility
|
|