KMID : 0357520140370030177
Journal of Radiological Science and Technology
2014 Volume.37 No. 3 p.177 ~ p.185
Effect of Automatic Exposure Control Marker with Chest Radiography in Radiation Reduction
Jung Ji-Sang

Choi Byoung-Wook
Kim Sung-Ho
Kim Young-Mo
Shim Ji-Na
Ahn Ho-Sik
Jin Duk-Eun
Lee Jae-Sik
Kang Sung-Ho
Abstract
This study focused on effects of patient exposure dose reduction with AEC (Auto Exposure Control) marker that is designed for showing location of AEC in X-ray Chest radiography. It included 880 adults who have to use Chest X-ray Digital Radiography system (DRS, LISTEM, Korea). AEC (Ion chambers are posited in top of both sides) are used to every adult and set X-ray system as Field size $17{ imes}17inch$ ¼ö½Ä À̹ÌÁö, 120kVp, FFD 180cm. 440 people of control group are posited on detector to include both sides of lung field and the other 440 people of experimental group are set to contact their lung directly to Ion chamber (making marker to shows location). Then, measured every DAP and, estimated patient effective dose by using PCXMC 2.0. The average age of control group (M:F=245:195) is 53.9 and the average BMI is 23.4. BMI ranges from under weight: 35, normal range: 279, over weight: 106 to obese: 20 and average DAP is 223.56mGycm2, Mean effective dose is 0.045mSv. The average age of experimental group (M:F=197:243) is 53.7 and the average BMI is 22.7. BMI ranges from under weight: 34, normal range: 315, over weight: 85 to obese: 6 and average DAP is 207.36mGycm2, Mean effective dose is 0.041mSv. Experimental group shows less Mean effective dose as 0.004mSv (9.7%) than control group. Also, patient numbers who got over exposure more than 0.056mSv (limit point to know efficiency of AEC marker) is 65 in control group (14.7%), 19 in experimental group (4.3%) and take statistics with t-Test. The statistical difference between two groups is 0.006. In order to use proper amount of X-ray in auto exposure controlled chest X-ray system, matching location between ion chamber and body part is needed, and using AEC marker (designed for showing location of ion chamber) is a way to reduce unnecessary patient exposure dose.
KEYWORD
AEC marker, Chest PA, Effective dose, Ion chamber, DAP
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)