A Study on the Apron Shielding Ratio According to Electromagnetic Radiation Energy
|
|
Jang Dong-Gun
Lee Sang-Ho Choi Hyung-Seok Son Ju-Cheol Yoon Chang-Yong Ji Yung-Sik Cho Yong-In Lee Hong-je Yang Seoung-Oh
|
|
Abstract
|
|
|
The medical institution has been used electromagnetic radiation of various energy. But researchers are divided on whether using apron for radiation shielding will be effective or not. The purpose of present study was to analyze electromagnetic radiation shielding effect of apron by using Monte carlo simulation. 1 MBq electromagnetic radiation was emitted from 10-500 keV at 10 keV increments in Monte carlo simulation. Then shielded radiation dose difference was confirmed, when 0.25 mmPb shield use for shielding. As a results, shielding ratio was markedly decreased in high energy electromagnetic radiation. The radiation dose was inversely increased with 0.25 mmPb shielding.
|
|
KEYWORD
|
|
Monte carlo simulation, shielding, electromagnetic radiation
|
|