KMID : 0357520150380020081
Journal of Radiological Science and Technology
2015 Volume.38 No. 2 p.81 ~ p.87
A Study on Compression Paddle Materials to Reduce Radiation Exposure Dose During Mammography; PC and PMMA and Carbon
Hong Dong-Hee

Jung Hong-Ryang
Abstract
This study was designed the band material in order to reduce the exposure pressure of the breast and the material was measured of Radiolucent and radiation properties with a radiation materials of PC, PMMA, Carbon. Also the image quality by image analysis to obtain the following results are below:
Unfors Xi dosimetry using radiation transmittance when the results of 8.353m§í is measured after re-moval of the cuff, the PC 6.308m§í, PMMA 6.223m§í, Carbon 7.218m§í were measured respectively. Semi-layer PC 0.375§®Al, PMMA 0.370§®Al, with Carbon 0.360§®Al Carbon, PC, PMMA was higher radiation properties and transparency in order.
nLight / OSL NanoDotTM dosimeter was used in the cuff and then removed by placing the dosimeter measured results center on 1.143m§í, at the edge 12.56m§í, on the central PC 8.990m§í, at the edge 10.291m§í, PMMA center on 8.391m§í , the edge 9.654m§í, on Carbon center 9.581m§í, 11.313 m§í were measured at the edge of Carbon, PC, PMMA showed a high permeability in order.
Image Pixel average J is then removed from the cuff 976.655, PC 831.032, PMMA 819.069, Carbon 897.118 Carbon, PC, PMMA was measured by high order.
KEYWORD
Mammography, Compression paddle, PC, PMMA, CARBON