KMID : 0357520150380020115
Journal of Radiological Science and Technology
2015 Volume.38 No. 2 p.115 ~ p.120
Improved Focal Liver Lesion Detection by Increasing Flip Angle During Gadoxetic Acid-Enhancement in MRI
Lee Se-Jy

Kim Young-Keun
Abstract
To study the differences of focal liver lesion image detection at 3 minute, 10 minute and 15 minute time points on gadoxetic acid (GA)¡¯s enhanced MR imaging with a flip angle (FA) of 30¡Æ compared with a 11¡Æ. The subjects were 69 patients evaluated with GA enhanced MR imaging with 3.0T MR scanner. The patients are total 35(23 men and 7 women at the mean age of 60.4 years), hepatocellular carcinoma(23) and metastsis(12) except for normal, cyst and hemangioma. After GA was injected, FA 11¢ª and 30¢ª images were obtained at 3 minute, 10 minute and 15 minute time points respectively. After quantitative and qualitative assessment of each image was done, statistical analysis was performed by using the independent sample T-test. From both quantitative and qualitative assessment of 3 minute and 10 minute MR images after the injection of GA, FA 30¡Æ images was found to be superior than FA 11¡Æ, but there were no statistical significance. However, at 15 minute time point, Statistically significant FA 30¡Æ image(p<0.05) was better than FA 11¡Æ therefore, the FA 30¡Æ improves the focal liver lesion detection. FA 30¡Æ of MR image can detect liver lesion more sensitively than the existing FA11¡Æ image after GA contrast enhancement at 15 minute time point.
KEYWORD
Magnetic resonance imaging, MRI, Gadoxetic acid, GA, Flip angle, FA, Focal liver lesion