Journal of Radiological Science and Technology
2016 Volume.39 No. 3 p.435 ~ p.441
Performance Evaluation of Domestic Prototype Dose Area Product Meter SFT-1
Lee Ho-Sun

Han Seong-Gyu
Roh Young-Hoon
Lim Hyun-Jong
Kim Jung-Min
Kim Jong-Uk
Chae Hyun-Sik
Yoon Yong-Su
Abstract
The importance of radiation dose display of medical X-ray equipment was emphasized, while third edition of IEC(International Electrotechnical Commission) 60601 started to apply. The existing medical X-ray equipment selected a method for attaching the DAP(Dose Area Product) meter when the dose display. However, because the DAP meter was dependent on all of the income, And it did not yet produced in Korea. So, we received the support of Seoul R&BD Program(Grants No. C1152055) to produce DAP meter prototype of the Domestically technology. In this study, the performance of this prototype was evaluated by comparing the German company's product Evaluation item was an electronic capture performance, radiation dose dependence, radiation quality dependence, energy transmittance, repeatability, light transmittance of 6 entries. And IEC 60580 was based on this evaluation. Evaluation results were electronic capture performance intrinsic error 9.5%, radiation dose dependence limits of variation 1%, repeatability coefficient of variation 2%, energy transmittance 91% each assessment was passed. However radiation quality dependence limits of variation 29%, light transmittance 55% was less than acceptance criteria.
KEYWORD
DAP meter, Domestic technology, Prototype, Performance evaluation