KMID : 0357520160390040607
Journal of Radiological Science and Technology
2016 Volume.39 No. 4 p.607 ~ p.614
Evaluation of DQA for Tomotherapy using 3D Volumetric Phantom
Lee Sang-Uk

Kim Jeong-Koo
Abstract
The study investigates the necessity of 3 dimensional dose distribution evaluation instead of point dose and 2 dimensional dose distribution evaluation. Treatment plans were generated on the RANDO phantom to measure the precise dose distribution of the treatment site 0.5, 1, 1.5, 2, 2.5, 3 cm with the prescribed dose; 1,200 cGy, 5 fractions. Gamma analysis (3%/3 mm, 2%/2 mm) of dose distribution was evaluated with gafchromic EBT2 film and ArcCHECK phantom. The average error of absolute dose was measured at 0.76¡¾0.59% and 1.37¡¾0.76% in cheese phantom and ArcCHECK phantom respectively. The average passing ratio for 3%/3 mm were 97.72¡¾0.02% and 99.26¡¾0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. The average passing ratio for 2%/2 mm were 94.21¡¾0.02% and 93.02¡¾0.01% in gafchromic EBT2 film and ArcCHECK phantom respectively. There was a more accurate dose distribution of 3D volume phantom than cheese phantom in patients DQA using tomotherapy. Therefor it should be evaluated simultaneously 3 dimensional dose evaluation on target and peripheral area in rotational radiotherapy such as tomotherapy.
KEYWORD
DQA , tomotherapy , 3D volume phantom , 3D dose evaluation