KMID : 0357520170400020245
Journal of Radiological Science and Technology
2017 Volume.40 No. 2 p.245 ~ p.251
Research of Protocols for Optimization of Exposure Dose in Abdominopelvic CT -
Hong Dong-Hee

Abstract
This study measured the exposure dose during abdominal-pelvic CT exam which occupies 70% of CT exam and tried to propose a protocol for optimized exposure dose in abdomen and pelvis without affecting the imagery interpretation. The study scanned abdomen-pelvis using the current clinical scan method, the 120 kVp, auto exposure control(AEC), as 1 phase. As for the newly proposed 2 phase scan method, the study divided into 1 phase abdomen exam and 2 phase pelvis exam and each conducted tube voltage 120 kVp, AEC for abdomen exam, and fixed tube current method in 120 kVp, 100, 150, 200, 250, 300, 350, 400 mA for pelvis exam. The exposure dose value was compared using $CTDI_{VOL}$ ¼ö½Ä À̹ÌÁö, DLP value measured during scan, and average value of CT attenuation coefficient, noise, SNR from each scan image were obtained to evaluate the image. As for the result, scanning of 2 phase showed significant difference compared to 1 phase. In $CTDI_{VOL}$ ¼ö½Ä À̹ÌÁö value, the 2 phase showed 26% decrease in abdomen, 1.8~59.5% decrease in pelvis for 100~250 mA, 12.7%~30% increase in pelvis for 300~400 mA. Also, DLP value showed 53% decrease in abdomen and 41~81% decrease in pelvis when scanned by 2 phase compared to 1 phase, but it was not statistically significant. As for the SNR, when scanning 2 phase close to heart, scanning 1 phase close to pelvis, scanning and scanning 1 phase at upper and lower abdomen, it was higher when scanning 2 phase for 200~250 mA. Also, the CT number and noise was overall similar, but the noise was high close to pelvis. However, when scanning 2 phase for 250 mA close to pelvis, the noise value came out similar to 1 phase, and did not show statistically significant difference. It seems when separating pelvis to scan in 250 mA rather than 400 mA in 1 phase as before, it is expected to have reduced effect of exposure dose without difference in the quality of image. Thus, for patients who often get abdominal-pelvic CT exam, fertile women or children, this study proposes 2 phase exam for smaller exposure dose with same image quality.
KEYWORD
Abdomino-pelvic CT exam , Exposure dose , Optimization , Protocol , Image evaluation
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)