Journal of Radiological Science and Technology
2017 Volume.40 No. 2 p.311 ~ p.316
Quantitative Analysis of Bone Mineral Measurements in Different Types of Dual-energy Absorptiometry Systems: Comparison of CT vs DEXA
Kim Myeong-Seong

Abstract
Generally assessing bone mineral density (BMD) were performed on dual energy X-ray absorptiometry (DEXA) the same as dual energy CT (DECT) with a rapid-kVp switching. The purpose of this study is to compare the different of BMD value between DEXA and DECT method, and evaluate usefulness of DECT method. Using scanner for BMD measurements were GE, Healthcare Discovery 750 HD for DECT and Hologic QDR 4500W for DEXA. For compare BMD value in each method, scanned lumbar spine phantom and subjects visiting Korean National Cancer Center from April 2015 to December 2015, records of 50 patients. This study was approved by the Institutional Review Board. The mean BMD value measures for spine phantom and for subjects in each scanners presented strong correlation (r=0.948 with p<0.05 for phantom; r=0.635 with p<0.05 and Kendall's tau (¥ó)=0.46 with p<0.05 for subjects) and linear relationship between DECT and conventional DEXA. DECT technique for BMD measurement will provide a very useful methodology without additional radiation
dose.
KEYWORD
DEXA , CT , BMD , DECT , osteoporosis