Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520170400030453
Journal of Radiological Science and Technology
2017 Volume.40 No. 3 p.453 ~ p.460
Development of Qualitative Evaluation of Medical Radiation Protective Apron
Lim Hei-Gyeom

Kim Sang-Hyun
Kim Tae-Hyung
So Woon-Young
Lee Tae-Hui
Kim Seung-Chul
Kim Jin-Tae
Abstract
This study proposes effective quality control and maintenance method by developing a new qualitative evaluation method of apron for medical radiation protection. As an experimental material, one of 0.45 mm lead and 100 of 0.45 mm Pb aprons were used and irradiated under the conditions of a tube voltage of 75 kVp and a tube current of 12.5 mAs to obtain an image. and using the Image J program, PSNR values were compared and analyzed. The results showed that there were 40 aprons (less than 11dB), 55 aprons (less than 11dB, less than 30dB), and 5 aprons (30dB or more). In addition, the dose showed a normal distribution for the apron, and 5 aprons with PSNR less than 11dB and 30dB or more were selected and divided into 8 zones, and these groups were statistically significant.
KEYWORD
Radiation protective apron , Lead , Image J , Pick signal-to-noise ratio , Quality control
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)