Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520180410010039
Journal of Radiological Science and Technology
2018 Volume.41 No. 1 p.39 ~ p.45
Evaluation of the Utility of a Volumetric Modulated Arc Therapy Specific Patient Quality Assurance using Software-based Quality Assurance System
Kang Dong-Jin

Jung Jae-Yong
Shin Young-Joo
Min Jung-Whan
Kim Yon-Lae
Kwon Kyung-Tae
Abstract
The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric M odulated A rc T herapy t reatment p lan. E valuate treatment plan t hrough t he D VH a nalysis, P TV m ean dose (ɤ????) and PTV 95% dose(ɤ ÉÝÉÙ ) compare the MFX based on original treatment plan, Average error rate was 0.9¡¾0.6%, 1.0¡¾0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9¡¾0.7%, 1.1¡¾0.7% (high dose region), 1.1¡¾0.9%, 1.2¡¾0.7% (low dose region). The average gamma though of MFX and Delta4PT is 98.7¡¾1.2%, 98.4¡¾1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.
KEYWORD
Software-based quality assurance, Log file, Gamma index, Patient specific quality assurance, Delivery quality assurance
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)