Journal of Radiological Science and Technology
2018 Volume.41 No. 5 p.445 ~ p.450
Evaluation of Radiation Doses of Dental Portable Equipment
Park Hoon-Hee

Kang Byung-Sam
Abstract
We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR
KEYWORD
Dental Leakage Dose, Dental Portable Equipment, Radiation Safety, Radiation Protection, Intra Oral Projection