Comparison of Noise Power Spectrum in Measurements by Using International Electro-technical Commission Standard Devices in Indirect Digital Radiography
|
|
Min Jung-Whan
Jeong Hoi-Woun Kim Ki-Won Kwon Kyung-Tae Jung Jae-Yong Son Jin-Hyun Kim Hyun-Soo
|
|
Abstract
|
|
|
The purpose of this study was to compare image quality of indirect digital radiography (IDR) system using the International Electro-technical Commission standard (IEC 62220-1), and to suggest the analysis of noise power spectrum (NPS) which were applied to IEC 62220-1 in medical imaging. In this study, Pixium 4600 (Trixell, France) which is indirect flat panel detector (FPD) was used. The size of image receptor (IR) is 7 ¡¿ 17 inch (matrix 3001 ¡¿ 3001) which performed 14bit processing and pixel pitch is 143 ¥ìm. In IEC standard, NPS evaluation were applied to RQA3, RQA5, RQA7 and RQA9. Because of different radiation quality, each region of interesting (ROI) were compared. The results of NPS indicated up to 3.5 mm-1 including low Nyquist frequency. RQA5 indicated the lowest NPS and the others indicated higher NPS results relatively. NPS result of ROI a38 was higher than ROI a92 and this result indicated that there are more noise in left (cathode) than right (anode). This study were to evaluate NPS by using different radiation quality and setting the each ROI, and to suggest the quantitative methods of measuring NPS
|
|
KEYWORD
|
|
Noise Power Spectrum, Indirect Flat Panel Detector, Region of Interest, Nyquist Frequency, Noise
|
|