Journal of Radiological Science and Technology
2018 Volume.41 No. 6 p.553 ~ p.560
Study on the Difference of Standardized Uptake Value in Fusion Image of Nuclear Medicine
Kim Jung-Soo

Park Chan-Rok
Abstract
PET-CT and PET-MRI which integrates CT using ionized radiation and MRI using phenomena of magnetic resonance are determined to have the limitation to apply the semi-quantitative index, standardized uptake value (SUV), with the same level due to the fundamental differences of image capturing principle and reorganization, hence, their correlations were analyzed to provide their clinical information. To 30 study subjects maintaining pre-treatment, 18F-FDG (5.18 MBq/§¸) was injected and they were scanned continuously without delaying time using BiographTM mMR 3T (Siemens, Munich) and Biograph mCT 64 (Siemens, Germany), which is an integral type, under the optimized condition except the structural differences of both scanners. Upon the measurement results of SUVmax setting volume region of interest with evenly distributed radioactive pharmaceuticals by captured images, SUVmax mean values of PET-CT and PET-MRI were 2.94¡¾0.55 and 2.45¡¾0.52, respectively, and the value of PET-MRI was measured lower by ?20.85¡¾7.26% than that of PET-CT. Also, there was a statistically significant difference in SUVs between two scanners (P<0.001), hence, SUV of PET-CT and PET-MRI cannot express the clinical meanings in the same level. Therefore, in case of the patients who undergo cross follow-up tests with PET-CT and PET-MRI, diagnostic information should be analyzed considering the conditions of SUV differences in both scanners.
KEYWORD
PET-CT, PET-MRI, Standardized Uptake Value, Attenuation Correction
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)