Evaluation of the Dose According to the Movement of Breath During Field-in-Field Technique Treatment of Breast Cancer Patients
Kwon Kyung-Tae

Abstract
Field-in-Field Technique is applied to the radiation therapy of breast cancer patients, and it is possible to compensate the difference in breast thickness and deliver uniform dose in the breast. However, there are several fields in the treatment field that result in a more complex dose delivery than a single field dose delivery. If the patient's respiration is irregular during the delivery of the dose by several fields and the change of respiration occurs, the dose distribution in the breast changes. Therefore, based on the computed tomography images of breast cancer patients, a human model was created by using a 3D printer (Builder Extreme 1000) to describe the volume in the same manner. A computerized tomography (CT) of the human body model was performed and a treatment plan of 260 cGy / fx was established using a 6-MV field-in-field technique using a computerized treatment planning system (Eclipse 13.6, Varian, USA). The distribution of the dose in the breast according to the change of the respiration was measured using a moving phantom at 0.1 cm, 0.3 cm, 0.5 cm amplitude, using a MOSOXIDE Silicon Field Effect Transistor (MOSFET, Best Medical, Canada) Were measured and compared. The distribution of dose in the breast according to the change of respiration showed similar value within ¡¾ 2% in the movement up to 0.3 cm compared to the treatment plan. In this experiment, we found that the dose distribution in the breast due to the change of respiration when the change of respiration was increased was not much different from the treatment plan.
KEYWORD
Field-in-Field Technique, Respiration variation, 3D print, Phantom dose, MOSFET
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)