Journal of Radiological Science and Technology
2019 Volume.42 No. 3 p.217 ~ p.222
The Effect of MLC Leaf Motion Constraints on Plan Quality and Delivery Accuracy in VMAT
Kim Yon-Lae

Chung Jin-Beom
Lee Jeong-Woo
Shin Young-Joo
Kang Dong-Jin
Jung Jae-Yong
Abstract
The purpose of this study is to evaluate the dose distribution by gantry rotation and MLC moving speed on treatment planning system(TPS) and linear accelerator. The dose analyzer phantom(Delta 4) was scanned by CT simulator for treatment planning. The planning target volumes(PTVs) of prostate and pancreas was prescribed 6,500 cGy, 5,000 cGy on VMAT(Volumetric Modulated Arc Therapy) by TPS while MLC speed changed. The analyzer phantom was irradiated linear accelerator using by planned parameters. Dose distribution of PTVs were evaluated by the homogeneity index, conformity index, dose volume histogram of organ at risk(rectum, bladder, spinal cord, kidney). And irradiated dose analysis were evaluated dose distribution and conformity by gamma index. The PTV dose of pancreas was 4,993 cGy during 0.1 §¯/deg leaf and gantry that was the most closest prescribed dose(5,000 cGy). The dose of spinal cord, left kidney, and right kidney were accessed the lowest during 0.1 §¯/deg, 1.5 §¯/deg, 0.3 §¯/deg. The PTV dose of prostate was 6,466 cGy during 0.1 §¯/deg leaf and gantry that was the most closest prescribed dose(6,500 cGy). The dose of bladder and rectum were accessed the lowest during 0.3 §¯/deg, 2.0 §¯/deg. For gamma index, pancreas and prostate were analyzed the lowest error 100% at 0.8, 1.0 §¯/deg and 99.6% at 0.3, 0.5 §¯/deg. We should used the optimal leaf speed according to the gantry rotation if the treatment cases are performed VMAT.
KEYWORD
VMAT, MLC, Homogeneity index, Conformity index, Gamma index
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)