KMID : 0357520190420050365
|
|
Journal of Radiological Science and Technology 2019 Volume.42 No. 5 p.365 ~ p.371
|
|
Performance Evaluation of the Developed Diagnostic Multi-Leaf Collimator and Implementation of Fusion Image of X-ray Image and Infrared Thermography Image
|
|
Kwon Soon-Mu
Shim Jae-Goo Chon Kwon-Su
|
|
Abstract
|
|
|
We have developed and applied a diagnostic Multi-Leaf Collimator (MLC) to optimized the X-ray field in medical imaging and the usefulness evaluated through the fusion of infrared image and X-ray image acquired by infrared camera. The hand and skull radiography with multi-leaf collimator(MLC) showed significant area dose reductions of 22.9% and 31.3% compared to ARC and leakage dose was compliant with KS A 4732. Also scattering doses of 50 cm and 100 cm showed a significant decrease to confirm the usefulness of MLC. It was confirmed that the fusion of infrared images with an adjustable degree of transparency was possible in the X-ray images. Therefore, fusion of anatomical information with physiological convergence is expected to contribute and improvement of diagnostic ability. In addition, the feasibility of convergence X-ray imaging and DITI devices and the possibility of driving MLC with infrared images were confirmed.
|
|
KEYWORD
|
|
Multi-Leaf Collimator, Medical Image Fusion, Digital Infrared Thermal Imaging, Scattered Radiation Dose, Dose Area Product
|
|