Journal of Radiological Science and Technology
2019 Volume.42 No. 6 p.483 ~ p.489
Comparison of Dose Measurement of Glass Dose Meter, Semiconductor Dose Meter, and Area Dose Meter in Diagnostic X-ray Energy
Son Jin-Hyun

Abstract
This paper obtained and compared these dose values by setting and comparing the X-ray imaging conditions (tube voltage 60 kVp, 70 kVp, 80 kVp, tube current 10 mAs, 16 mAs and X-ray field size are 10 ¡¿ 10 cm, 15 ¡¿ 15 cm). Each dose value was measure 10 times and represented as an average value. The purpose of this experiment is to serve as a reference for the X-ray exposure of diagnostic areas according to the type of dosimeter and to help with another dose measurement. The results of the experiment showed very little difference between the glass dosimeter(GD) and semiconductor dosimeter values due to changes in tube voltage of 60, 70, 80 kVp, regardless of field sized, but for dose area product(DAP), the difference in dose value was significant according to field size.
KEYWORD
Glass Dosimeter, Semiconductor Dosimeter, DAP, kVp, mAs, Field size