Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520200430040251
Journal of Radiological Science and Technology
2020 Volume.43 No. 4 p.251 ~ p.257
Study on the Resolution Characteristics by Using Magnetic Resonance Imaging 3.0T
Min Jung-Whan

Jeong Hoi-Woun
Han Ji-Hyun
Lee Si-Nae
Han Song-Yi
Kim Ki-Won
Kim Hyun-Soo
Son Jin-Hyun
Abstract
This study was purpose to quantitative evaluation of edge method of modulation transfer function(MTF) and physical image characteristics of by obtain the optimal edge image by using magnetic resonance imaging(MRI). The MRI equipment was used (MAGNETOM Vida 3.0T MRI, Siemense healthcare system, Germany) and the head/neck matrix shim MR coil were 20 channels(elements) receive coil. The MTF results of showed the best value of 0.294 based on the T2 Nyquist frequency of 1.0 mm-1. The MTF results of showed that the T1 image is 0.160, the T1 CE image is 0.250, T1 Conca2 image is 0.043, and the T1 CE (Concatenation) Conca2 image is 0.190. The T2 image highest quantitatively value for MTF. The physical image characteristics of this study were to that can be used efficiently of the MRI and to present the quantitative evaluation method and physical image characteristics of 3.0T MRI.
KEYWORD
Magnetic resonance imaging, Modulation transfer function, Edge method, Fujita method, Quality Assurance
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)