Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520200430050337
Journal of Radiological Science and Technology
2020 Volume.43 No. 5 p.337 ~ p.341
Assessment of Organ Dose in Mammoplasty Patient by Monte Carlo Simulation during Mammography
Kim Ji-Soo

Cho Yong-In
Kim Jung-Hoon
Abstract
Although the screening with a mammography has been shown to be economical, simple and effective in detecting breast cancer, it is accompanied by the risk from radiation. Therefore, this study analyzed the glandular dose and organ dose according to the target-filter combination and the presence and absence of implants using Monte Carlo simulation. The results indicate that at a tube voltage of 30 kV and a tube current of 50 mAs, the dose increased in the order of Mo/Mo. Mo/Rh, Rh/Rh and W/Rh in proportion to the atomic number of the target-filter. In addition, in phantom without implant a reduction in dose was seen when compared to the phantom with implant. The organ dose was highest in the lens except for the breast on the examination side regardless of the presence or absence of the implant. These results may contribute to use basic data for the diagnostic reference level of breast plastic surgery patients.
KEYWORD
Mammography, Breast implant, Organ dose, Monte carlo simulation, Breast dose
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)