KMID : 0357520220450010011
Journal of Radiological Science and Technology
2022 Volume.45 No. 1 p.11 ~ p.17
Evaluation of the Resolution Characteristics by Using American College of Radiology Phantom for Magnetic Resonance Imaging
Min Jung-Whan

Jeong Hoi-Woun
Han Ji-Hyun
Lee Si-Nae
Kim Min-Je
Kim Seung-Chul
Abstract
This study was purpose to quantitative assessment of the resolution characteristics by using American college of radiology(ACR) phantom for magnetic resonance imaging (MRI). The MRI equipment was used (Achiva 3.0T MRI, Philips system, Netherlands) and the head/neck matrix shim SENSE head coil were 32 channels(elements) receive MR coil. And the MRI equipment was used (Discovery MR 750, 3.0T MRI, GE medical system, America) and the head/neck matrix shim MC 3003G-32R 32-CH head coil were receive MR coil. As for the modulation transfer function(MTF) comparison result by using ACR magnetic resonance imaging phantom, the MTF value of the ACR standard T2 image in GE equipment is 0.199 when the frequency is 1.0 mm-1 and the MTF value of the hospital T2 image in Philips equipment is 0.528. It was used efficiently by using a general sequence more than the standard sequence method using the ACR phantom. In addition it is significant that the quantitative quality assurance evaluation method for resolution characteristics was applied mutatis mutandis, and the result values of the physical image characteristics of the 3.0T MRI device were presented.
KEYWORD
ACR Magnetic resonance imaging Phantom, Modulation transfer function, Edge method, Fujita method, Quality Assurance