Development of the Upper Wear Fixation Device for Chest AP X-ray Imaging on the Emergency Stretcher Bed
Lim Woo-Taek

Hong Dong-Hee
Abstract
This study aimed to provide basic data for 3D printing in the medical health field by developing upper wear fixation device (UWFD), an auxiliary device for shortening chest AP examination time on emergency room beds and non-contact with patients. The standard of hooks was modeled according to the bed frame using the Autodesk Fusion 360. It was printed with Form2 (Formlabs, Somerville, MA, USA), as SLA (stereo lithography apparatus) method, and was washed and hardened using Form Wash and Form Cure. The completed UWFD conducted an online survey on 4 items of stability, convenience, availability, preference and general characteristics. The total stability average was 3.93¡¾0.80, the total convenience average was 3.93¡¾0.68, the total availability average was 4.01¡¾0.89, and the total preference average was 3.80¡¾1.08. This study was significant in suggesting improvements in the general X-ray examination process in the emergency room by designing and making aids to easily fixing the patient's top to the frame of the emergency bed while meeting promptness and non-contact with the patient.
KEYWORD
Emergency Patient, Chest X-ray, Upper wear fixation device, Emergency Bed, Contact
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)