Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357520230460040295
Journal of Radiological Science and Technology
2023 Volume.46 No. 4 p.295 ~ p.301
A Study on Characteristic of Image Quality according to CT Table Height in Computed Tomography
Kim Ki-Won

Min Jung-Whan
Lee Sang-Sun
Lee Young-Bong
Lee Ki-Jong
Park Han-Sol
Oh Joo-Young
Abstract
In addition to protocol adjustments during CT examinations, the height of the CT table can also affect image quality. Therefore, this study aimed to investigate the change in image quality depending on the height of the table in brain CT, which accounts for a large proportion of CT examinations, by measuring signal to contrast to noise ratio (CNR) and noise power spectrum (NPS) using the head phantom and evaluating them. The head phantom images were acquired using Philips Brilliance iCT 256. When the image was acquired, the table height was adjusted to 815, 865, 915, 965, 1015, and 1030 mm, respectively, and each scan was performed 3 times for each height. The CNR result showed the highest value at 965 mm, which is the height adjacent to the center of the head phantom. NPS showed the lowest NPS at 915 mm, the center of the head phantom in the low frequency region. From these results, it can be seen that the height of the table in CT examination is closely related to the image quality, and it can be seen the characteristics of image quality according to CT table through quantitative evaluation methods such as CNR and NPS.
KEYWORD
Computed Tomography, CT Table, Image Quality, Contrast to Noise Ratio, Noise Power Spectrum
FullTexts / Linksout information
Listed journal information