Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0357919750090020213
Korean Journal of Pathology
1975 Volume.9 No. 2 p.213 ~ p.220
A Comparative Study on Methods for the Dtermination of SGOT-Activities
ÃÖÀμº/In-Seong Choe
Àå¹®±â/¹®¿µÈñ/¹é½Â·æ/Moon-Ki Chang/Young-Hoo Moon/Sung-Yong Paik
Abstract
In order to compare the accuracy of two methods, conventional Reitman-Frankel
method and single vial enzyme test kit method, for the determination of serum
glutamic-oxalacetic tran-saminase (SGOT) activities, total 111 serum samples were
examined with reference tests of revised Henry method. In the case of SGOT activity
being less than 50 A.U., the two test methods and the reference method were in good
agreement, respectively. However, in the range of SGOT activity over 50 A.U., which is
very significant in diaguostic evalution, the enzyme kit method excellently agreed with
the reference method, whereas the value of SGOT activity determined by
Reitman-Frankel method was significantly lower than by the references test. It is
suggested that the method of enzyme test kit should be used for the determination of
SGOT in clinical diagnosis.