Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0359020130460010095
Korean Journal of Gastrointestinal Endoscopy
2013 Volume.46 No. 1 p.95 ~ p.97
Fracture of Self-Expandable Metal Stent during Endoscopic Removal in Benign Biliary Stricture
Joo Kyu-Re

Paik Chang-Nyol
Chung Woo-Chul
Lee Kang-Moon
Yang Jin-Mo
Abstract
The endoscopic insertion of the self-expandable metal stent (SEMS) in benign biliary stricture has become an alternative to surgery. Fracture or migration of SEMS can occur rarely as complications. We report a case of fracture of SEMS during endoscopic retrieval in patients with chronic pancreatitis. In this case, broken stent was successfully removed with endoscopic ballooning of bile duct and with a snare device.
KEYWORD
Self-expandable metal stent, Fracture, Benign biliary stricture