KMID : 0359620150270030181
|
|
Journal of Korean Knee Society 2015 Volume.27 No. 3 p.181 ~ p.186
|
|
Biomechanical Study of the Fixation Plates For Opening Wedge High Tibial Osteotomy
|
|
Kim Kug-Jin
Song Eun-Kyoo Seon Jong-Keun Seol Jong-Hwan
|
|
Abstract
|
|
|
Purpose: The purpose of this study was to compare the mechanical stability of three types of plate systems for opening wedge high tibial osteotomy.
Materials and Methods: Forty-eight fresh frozen porcine tibia specimens were assigned to three different fixation device groups: Aescular group (16 specimens) was fixed with Aescular plates; Puddu group (16 specimens) with a Puddu plate, and TomoFix group (16 specimens) with a TomoFix plate. We compared axial displacements under compression loads from 200 to 2,000 N and maximal loads at failure among 8 specimens per group. We also compared displacements under cyclic load after 100 cycles at a compressive load of 2,000 N among 8 specimens per group.
Results: In all three groups, displacement under compression load increased with the increase in the axial compressive load; however, no significant intergroup differences were observed in the mean values under tested loading conditions. The mean maximal loads at failure were not significantly different (6,055, 6,798, and 6,973 N in the Aescular, Puddu, and TomoFix groups, respectively; p=0.41). While the TomoFix group showed less extension and strain during the cyclic load test, the mean values showed no significant differences among groups.
Conclusions: All three plate systems were found to provide fixation stability suitable for bearing axial compression and cyclic loads while walking.
|
|
KEYWORD
|
|
Knee, High tibial osteotomy, Opening wedge, Fixation plate, Biomechanical study
|
|