KMID : 0359920070260040448
Korean Journal of Nephrology
2007 Volume.26 No. 4 p.448 ~ p.454
Glucose Pump Technique is as Good as Ultrasound Dilution Technique for Vascular Access Surveillance in Hemodialysis Patients
Ha Sang-Jin

Cho Byung-Hyun
Lee Yoon-Jung
Chung Kyung-Hwan
Moon Ju-Young
Lee Sang-Ho
Lee Tae-Won
Ihm Chun-Gyoo
Abstract
Purpose: Vascular access flow (Qa) measurements are important in the surveillance protocol of hemodialysis vascular access stenosis. The glucose pump technique (GPT) is a technique for Qa measurement based on the dilution technique of a constant glucose infusion that was introduced in 1995. The aim of this study is to verify the clinical efficacy of GPT in vascular access surveillance compared with ultrasound dilution technique.

Methods: In 31 chronic hemodialysis patients with a AV fistula, we compared Qa measurements performed with GPT in pre-dialysis and the ultrasound dilution technique (Transonic HD01, Transonic System Inc., Ithaca, NY) during hemodialysis.

Results: Mean Qa was 1171 ml/min by GPT versus 1028 ml/min by HD01 (p=0.262). There was a strong linear correlation between the two methods (r=0.61; p<0.01). Ultrasound dilution technique HD01 yielded 6 cases of high risk Qa measurements and GPT yielded 4 cases of high risk Qa measurements. The diagnostic accuracy of GPT tested with the ROC curve was similar with ultrasound dilution technique HD01. The specificity and sensitivity was 80% and 66.7% according to Qa=600 ml/min. But sensitivity was 83.3% according to Qa=750 ml/min.

Conclusions: GPT offers the advantage of a simple and economic bedside procedure easily performed before dialysis and had a similar diagnostic accuracy and efficiency compared with ultrasound dilution technique HD01.
KEYWORD
Stenosis, Hemodialysis, Dilution technique