KMID : 0360220120530050662
Journal of the Korean Ophthalmological Society
2012 Volume.53 No. 5 p.662 ~ p.667
Intraocular Pressure Measurement with the Noncontact Tonometer and Rebound Tonometer through Plano Soft Contact Lenses
Lee Chang-Mok

Yoo Young-Cheol
Abstract
Purpose: To investigate the accuracy of intraocular pressure (IOP) measurement through a plano soft contact lens (SCL) in situ by a noncontact tonometer (NCT) and a rebound tonometer (RBT).

Methods: The IOP of 66 eyes of 33 subjects with no ocular pathology was measured by NCT (tonometer CT-80, TOPCON, Japan) and RBT (iCare rebound tonometryTM, Tiolat Oy, Finland). IOP measurement was repeated through plano SCL(ACUVUE OASYS¢ç, Johnson & Johnson Vision Care Inc., USA) in situ. Statistical analysis was performed using the paired t-test.

Result: Mean IOP measured by NCT was 13.10 ¡¾ 2.52 mmHg without SCL and 12.95 ¡¾ 2.56 mmHg with SCL. Mean IOP measured by RBT was 14.13 ¡¾ 2.94 mmHg without SCL and 13.84 ¡¾ 2.75 mmHg with SCL. No significant statistical differences were found between IOP measured with and without SCL (p=0.47, p=0.11 respectively).

Conclusions: Reliable measurement of IOP by NCT and RBT can be achieved through plano SCL.
KEYWORD
Intraocular pressure, Noncontact tonometer, Rebound tonometer, Soft contact lens