KMID : 0360220140550070984
Journal of the Korean Ophthalmological Society
2014 Volume.55 No. 7 p.984 ~ p.990
Comparative Analysis of Corneal Refractive Power Measured with AL?Scan¢ç, Autokeratometer, and Pentacam¢ç
Park Sung-Jin

Lim Sung-Hyup
Lee Ho-Young
Abstract
Purpose: To investigate clinical availability of AL-ScanTM (Nidek, Gamagori, Japan) by comparing corneal refractive power with AL-ScanTM, AutokeratometerTM (Topcon KR-1, Tokyo, Japan) and PentacamTM (Oculus, Wetzlar, Germany) devices.

Methods: Seventy-one patients (142 eyes) who visited our hospital for refractive surgery were tested using AL-Scan¢ç, Autokeratometer and Pentacam¢ç and corneal refractive power was compared among devices.

Results: When comparing measurements with AL-Scan¢ç, Autokeratometer and Pentacam¢ç, the mean corneal refractive power was 43.37 ¡¾ 1.32 D (2.4 mm zone), 43.35 ¡¾ 1.32 D (3.3 mm zone), 43.36 ¡¾ 1.35 D, and 43.35 ¡¾ 1.36 D respectively and showed no significant differences. Corneal refractive power had strongly positive linear correlation (p < 0.001) and Bland-Altman plots showed high degree of agreement among AL-Scan¢ç, Autokeratometer and Pentacam¢ç devices.

Conclusions: Because measuring ocular biometry with AL-Scan¢ç including axial length, intraocular lens power calculation and topography simultaneously is possible, clinical use is convenient. Corneal refractive power was not different when compared with autokeratometer and Pentacam¢ç devices, thus, AL-Scan¢ç can be used in the clinical environment.
KEYWORD
AL-Scan¢ç, Autokeratometer, Corneal refractive power, Pentacam¢ç