KMID : 0360220150560050694
Journal of the Korean Ophthalmological Society
2015 Volume.56 No. 5 p.694 ~ p.701
Comparison of Central Corneal Thickness and Anterior Chamber Depth Measured Using Three Different Devices
Han Se-Hoon

Hwang Ho-Sik
Shin Min-Chul
Han Kyung-Eun
Abstract
Purpose: To compare measurements of central corneal thickness (CCT) and anterior chamber depth (ACD) obtained using Galilei¢â, Pentacam¨Þ (Oculus, Wetzlar, Germany) and Lenstar¨Þ (Haag-Streit, Koeniz, Switzerland) and analyze the measurement agreements.

Methods: CCT and ACD were measured using Galilei¢â, Pentacam¨Þ and Lenstar¨Þ in 47 eyes of 25 healthy subjects. The measurements were compared among the 3 devices.

Results: The average CCT measurements using Galilei¢â, Pentacam¨Þ and Lenstar¨Þ were 552.6 ¡¾ 29.41 ¥ìm, 543.9 ¡¾ 30.50 ¥ìm and 537.5 ¡¾ 30.26 ¥ìm, respectively. The measurements significantly correlated with each other (r > 0.9, p < 0.001), but were statistically significantly different (p < 0.001). The average ACD measurements using Galilei¢â, Pentacam¨Þ and Lenstar¨Þ were 3.23 ¡¾ 0.360 mm, 3.22 ¡¾ 0.403 mm and 3.19 ¡¾ 0.367 mm, respectively. The measurements significantly correlated with each other (r > 0.9, p < 0.001), but were statistically significantly different (p = 0.034). The CCT 95% limits of agreement (LoA) between Galilei¢â and Pentacam¨Þ, Pentacam¨Þ and Lenstar¨Þ and Lenstar¨Þ and Galilei¢â were 31.95 ¥ìm, 44.76 ¥ìm and 46.57 ¥ìm, respectively and 95% ACD LoA were 0.46 mm, 0.32 mm and 0.28 mm, respectively.

Conclusions: CCT and ACD measured using the 3 devices were highly correlated with each other but the measurements were statistically different. Therefore, the measurements were not interchangeable and these differences should be considered in clinical use.
KEYWORD
Anterior chamber depth, Central corneal thickness, Galilei¢â, Lenstar?, Pentacam?